7 取样方法
1 空气 2 样品流
设定旁通流速
对于多样品流,切勿将螺钉拧松 4 圈以上。
旁通流量可以在分析仪处于关机模式时进行调整。按 8 9 所示,利用流量阀调整采样旁通管线
的流速。请参阅 规格 106 了解样品流速范围。利用一个外部流量计测量采样旁通管线的流速。当
样品流远离分析仪时,增加采样旁通管线的流速,以便更快地对样品流的变化作出响应。
8 旁通流速调整 单流
中文
97

Table of Contents

Other manuals for Hach 5500sc SiO2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 5500sc SiO2 and is the answer not in the manual?

Hach 5500sc SiO2 Specifications

General IconGeneral
Power Requirements100 - 240 VAC, 50/60 Hz
Resolution0.001 mg/L
Light SourceLED
DetectorPhotodiode
Sample Flow Rate200 to 500 mL/min
Operating Temperature5 to 45 °C
Sample Temperature5 to 45 °C
Outputs4-20 mA
EnclosureNEMA 4X/IP66
MethodHeteropoly blue method