Functions and outputs
184 I3031-14.0 HBM: public Quantum
X
Peak value function
Keep in mind the following notes when working with peak
value monitoring channels:
S The maximum sampling rate is 4800 Hz
S The sampling rate of the channel for peak value moni
toring must not be higher than the sampling rate of
the input channel.
S The peak value units do not allow for any other peak
value units or RMS values as input
Each peak value detection unit can monitor either the
Min peak or the Max peak of one of the four signals
within the system identified as "isochronous".
A peak value unit may operate in different modes:
S EXPORT: peak values will be continuously updated
S HOLD: last peak value will be "frozen"
S PEAK VALUE: peak value detection enabled
S FOLLOW: peak value detection disabled, i.e. the
channel returns the original signal of the input channel
The following combinations are possible:
RUN MAX PEAK VALUE
HOLD MAX PEAK VALUE
FOLLOW HOLD MAX
This also applies to minima.