7.2.10 Output measurement signals to CAN bus
(MX840B)
The MX840B amplifier allows channels 28 to output to
the CANbus (channel 1). This mode is configured entirely
in the MX Assistant.
Fig. 7.9 Output to CAN bus (MX840A, connection 1)
7.2.11 Output measurement signals to CAN bus
(MX471B)
The MX471B module allows measurement signals, or the
signals calculated in real time, to be output to the CAN
Bus. This gateway mode is typically used in test benches
or in mobile measuring mode, for connection to a central
CANbased data logger.
This mode is configured entirely in the MX Assistant soft
ware. The signals to be transmitted must be parameter
ized isochronously (in real time), and then assigned to
the relevant CAN port. The parameterization is perman
ently stored in the modules (EEPROM). To simplify integ
ration at the opposite end (e.g. logger/test bench), the
MX Assistant can generate a CAN database of signals
(*.dbc).