Connecting individual QuantumX modules
Quantum
X
I3031-14.0 HBM: public 83
that connect network chains together in star configura
tions.
A hop is the transition from one module to another (this
means n-1 hops for n QuantumX modules in a chain).
Depending on the connection situation, 1 to 2 hops are
counted in one hub (see Fig. 7.14).
To count the total number of hops, the longest chain to
the data sink must be counted (worst case).
123
1 234
67891011
Hub
Displayed:
Total number of modules: 15
Total number of hops: 12
QuantumX module
Longest chain to data sink (max. 12)
Data sink
P3
P4
P3 = Port 3
12
5
Fig. 7.13 Example of a star topology with two chains and one
hub