Modules and transducers
Quantum
X
I3031-14.0 HBM: public 95
Background calibration is available for the following
amplifiers: MX840B, MX440B, MX1615B, MX430B and
MX238B.
These amplifiers have a second measurement circuit in
the full/half bridge measurement mode, which measures
in parallel to the input circuit and implements a calibration
cycle in a 30 second rhythm. This ensures long-term and
short-term stability in the circuit. The accuracy of the cal
ibration channel is then transferred to the measurement
channel with a patented process.
These channels therefore demonstrate high stability with
respect to self-heating.
Background calibration can be parameterized with the
QuantumX Assistant or using catmanEASY®.
Background calibration is switched on the default set
tings. The cyclical calibration can be parameterized via
the QuantumX Assistants and via catmanEASY®.