Command reference Menu 6. SETTINGS
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Error 13 from
This value specifies from which Pascal value the device should report error signal 13.
The value can be in the range from -999 … 0 Pascal.
ATTENTION!
The error limit F13 is defined as -10 Pa by default. Any other value
could result in faulty measurements. This value should therefore only
be changed in consultation with HEMATECH PRÜFTECHNIK.
Decimal places Pa
Here you can select the number of decimal places of measurement unit Pa. This value
determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 2.
The default value is 0
Decimal places mbar
Here you can select the number of decimal places of measurement unit mbar. This value
determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 3.
The default value is 0
Decimal places psi
Here you can select the number of decimal places of measurement unit psi. This value
determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 4.
The default value is 4
Decimal places mmWC
Here you can select the number of decimal places of measurement unit mmWS. This
value determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 1.
The default value is 1
Decimal places mmHg
Here you can select the number of decimal places of measurement unit Hg. This value
determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 2.
The default value is 2
Decimal places ml/min
Here you can select the number of decimal places of measurement unit ml/min. This val-
ue determines in what number of decimal places the measurement unit is displayed.
The range is 0 ... 3.
The default value is 1
Make sure you have made all necessary settings in this dialog window.
Turn the key switch anti-clockwise again to switch the write protection ON.
An audible signal will sound and the device saves the changed data to the
RAM memory.
Press the ESC-key to exit the menu.