Command reference Menu 1. AUTOMATIC
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Sequencer
The sequencer function can be globally activated or deactivated.
(see sequencer setup)
With the sequencer activated, the transmitted program number is interpreted as Se-
quencer Number. The transmitted program number must only be in the range from 1
to 10. Otherwise an error message 3 (Program not available) will be submitted.
With the Start signal the steps memorized under the selected sequencer number are
successively processed.
Besides the program number the dialog window Automatic also shows the sequencer
number Seq: -- and the active step Step: --.
With the sequencer activated the number of passed cycles and the total number of cy-
cles are displayed in between the measuring result and the measuring value unit.
Dialog window AUTOMATIC SEQUENCER
If an error occurs in the sequence, the output error will be set and the complete test
sequence aborted. The error must be acknowledged as specified for the corresponding
machine interface. After a restart the complete test sequence is repeated.
If the input Abort is set, the entire test sequence will also be aborted. A restart will thus
also start with the first activated test step.
In dependence on the setting of the switch NIO=continue for the corresponding test step,
the sequence is aborted or continued in case of a measuring result unequal I.O..
At the end of any test step the output of the corresponding test result is set for a short
time.
If none of the 10 possible steps is activated in the selected sequencer, the error mes-
sage no step activated with error number 40 will be generated.