Maintenance Errors, Causes and Remedy
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
109
Errors, Causes and Remedy
The test device monitors valves, measuring transducers, specimen and the function se-
quence. In case of an error, this will be indicated by the measuring value display and
printed out as error code via the connected printed, in Automatic mode a floating con-
tact
Fault
will additionally be switched. A visual message
Error
will be emitted by light
indicators.
Acknowledging error messages
Depending on the operation mode the error messages must be acknowledged in dif-
ferent ways:
Operation mode MANUAL
After eliminating the error press the ESC-key to return to the dialog window
MANUAL MENU,
or
press the ENTER-key to repeat the selected function, after the error has been
corrected.
Operation mode AUTOMATIC (Auto)
After rectifying the error remove the signal AUTOMATIC (X 21/6).
Operation mode AUTOMATIC (SemiAuto)
After rectification of the error apply the START-signal to interface (X 21/2),
or
press the ESC-key
Operation mode AUTOMATIC (Minimal)
After rectification of the error apply the START-signal to interface (X 21/2) again,
or
press the ESC-key.