Performance characteristics Product description
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Performance characteristics
The test device offers the user a number of functions and possibilities to set parameters
for measurements, as well as for the output and distribution of these measuring results:
Function monitoring
The device takes over the monitoring of parts and function sequence. The user re-
ceives error messages through the plain text display and floating contact.
99 Test programs
There is a possibility to create up to 99 different programs and to select these either
via the control panel, or by means of external signals (BCD, 8 bit) in automatic opera-
tion.
Data exchange
All test parameters and measuring results can be entered and output, printed or pro-
cessed further by PC through the serial interface.
Result signals
The device delivers several result signals:
− I.O. (Accept)
− Rework
− Reject and gross leak (gross leak message only through V24 interface)
− Envelope curve,
Filling fault. (optional)
− Fault
The result signals can vary depending on device variant or have different designations.
Fill pressure, underfill pressure, underfill time and test pressure,
are separately set/disabled via keypad for each test program.
Times for fill, stabilization, pause, tare, measure and vent,
are separately adjustable via keypad from 0.1 … 999.9 seconds for each test pro-
gram.
Measuring range
∆ p 0 … 3000 Pa.
Resolution 1 Pa; limit contact setting up to max. 2000 Pa
(100 Pa = 1 mbar = 10 mmWC).
Measuring value display
You can switch over to:
PASCAL, mbar, psi, mmWC, mmHg, ml/min