Command reference Menu 2. MANUAL
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
39
1.1 Determination of a reference record
Connect a leak free original work piece to the tester.
Proceed as follows to determine the reference curve:
enter the number of the desired program (in this example: 1) into the top line and
confirm the entry with the ENTER-key.
Turn the key switch clockwise to horizontal position (input mode ON).
Press key F2 to start the test without leak simulation.
You can follow the test sequence by the horizontal progress bar, the help bar addi-
tionally shows the corresponding actual status of the test sequence.
Dialog window for manual testing
Once the test is completed, the display will show the following message.
Dialog window manual test, determination of reference curve
Turn the key switch anti-clockwise to vertical position to save the reference curve
you have determined. An audible signal will sound and the determined data will
be written to the RAM memory.
Press ESC to determine a new reference curve.
A new measurement will be executed.