Trouble Shooting Chart Maintenance
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
Filling pressure
after expiration of
filling time too low
Check the network pressure
(see chapter
Connecting the
test medium hose
Pressure system
outside tolerance
Redetermine the pressure
system in menu 6. Settings/3.
Pressure system
Filling pressure
after expiration of
filling time too high
Pressure system
outside tolerance
Redetermine the pressure
system in menu 6. Settings/3.
Pressure system
Filling pressure
after expiration of
stabilization time
too low
Network pressure too
low, coarse leak on
specimen,
Check the network pressure
(see chapter
Connecting the
test medium
)
Pressure system outsi-
de tolerance
Redetermine the pressure sys-
tem in menu 6. Settings/3.
Pressure system
Stabilization time
too short
Check the stabilization time
Filling pressure
after expiration of
stabilization time
too high
Pressure system
outside tolerance
Redetermine the pressure sys-
tem in menu 6. Settings/3.
Pressure system
Replace the middle section
of the shut-off valve*
Pressure system not
adjustable
Check the network pressure
(see chapter
Connecting the
test medium
)
Test medium connec-
tion not connected
Connect the test medium con-
nection
Pressure system
defective
Inform the customer service
Output board
X5 defective
Replace the output board*
Replace the P/I-converter*
Rated pressure out
of range
Network pressure too
low during pressure
determination
Redetermine the pressure sys-
tem in menu 6. Settings/3.
Pressure system