Command reference Menu 3. SETUP
EN 3925-0060 V207-00 MANUAL R03.doc
HeMaTech
Prüftechnik
75
Examples:
Single enquiry of the status of test pressure valve "Y2" on
leak tester 3925-0050:
Command
Feedback
DOU1<CR>
0<CR>
Or
1<CR>
Multiple enquiry of the status of test pressure valve "Y2" and shut-off valve "Y3" on
leak tester 3925-0050:
Command
Feedback
DOU1-2<CR>
0,0<CR>
Or
Or
DOU1;2<CR>
0,1<CR>
Or
1,0<CR>
Or
1,1<CR>
Single setting of the status of test pressure valve "Y2" on
leak tester 3925-0050:
Command
DOU1=1<CR>
Multiple setting of the status of test pressure valve "Y2" and shut-off valve "Y3" on
leak tester 3925-0050:
Command
DOU1-2=1;1<CR>
Or
DOU1;2=1;!<CR>
Further information concerning the command reference can be found in the command
reference manual for the leak tester (PLC95).