121
_____________________________________________________________________
7.3 Setting the Channel Screen
______________________________________________________________
Flashing cursor
GND
Grounds the input (to check zero position).
DC
DC coupling (DC and AC components of input
signal)
CLAMP
Use when a 9018-10, 9132-10 CLAMP ON PROBE
is connected.
NOTE
When measuring current using a HIOKI 9018-10, 9132-10
CLAMP ON PROBE, setting the input coupling to CLAMP
automatically selects the A (Amperes current) display. When this
occurs, set the current range of the probe and that of the product
to the same range.
The zero position may be offset when the input coupling is DC
or GND and the input is open circuit in the high-sensitivity (10
mV) range.
7.3.4 Setting the Input Coupling
The input coupling for the input signal is set.
Procedure
Channel Screen
1. Press the
SET
button to display the Channel Screen.
2. Move the flashing cursor to the desired channel position.
3. Use the
buttons to make a setting.