23
_____________________________________________________________________
2.10 Notes on Measurement
______________________________________________________________
DANGE
R
To avoid the risk of electric shock and damage to the
product
Do not apply more than 450 V AC or DC between the
analog inputs and product, or between the analog
inputs.
Terminal inputs should never exceed the limits
indicated below.
Input/output terminals Maximum input voltage
Analog input 450 V AC rms, DC
EXT TRIG -5 V to +10 V
TRIG OUT -20Vto+30V
500 mA max, 200 mW max
The maximum voltage applied to the probe and cables
connecting to the analog input of the product must
never exceed the limits in the following table.
Notwithstanding these limits, the voltage at the analog
input must not exceed 450V AC or DC.
Maximum input voltage
9197
CONNECTION CORD
500 V AC rms, DC
9217
CONNECTION CORD
300 V AC rms, DC
9322
DIFFERENTIAL PROBE
1000 V AC, 2000 V DC (CAT ll)
600 V AC, DC (CAT lll)
2.10 Notes on Measurement