38
_____________________________________________________________________
4.3 Setting Items
______________________________________________________________
Flashing cursor
Using channels Constant recording length [DIV]
1 channel 20, 50, 100, 200, 400, 800, 1600, 3200
1 and 2 channels 20, 50, 100, 200, 400, 800, 1600
1 to 4 channels
(8808-01, 8808-51
only)
20, 50, 100, 200, 400, 800
4.3.3 Setting the Recording Length
The length of recording for one measurement operation
(number of divisions) can be set.
Constant recording length mode
: Select the recording length.
Variable recording length mode
: Variable recording length can
be selected by the user.
Status Screen
Procedure 1 Constant recording length mode
1. Press the
SET
button to display the Status Screen.
2. Move the flashing cursor to
Shot
and select the recording
length using the
buttons.
Depending on the number of channels to be used, the
recording length is limited. (For setting the number of
channels, refer to
Section 4.8.1
.)