141
_____________________________________________________________________
8.4 Pre-trigger
______________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
IND
-95
50% 50%
95%
Trigger point
Pre-trigger: 95%
Pre-trigger: 50%
Pre-trigger: -95%
10 DIV
5 DIV 20 DIV
15 DIV
hen the recordin
len
th is 25 DIV
Trigger point
Pre-trigger: 0 DIV
Pre-trigger: 5 DIV
Pre-trigger: 10 DIV
25 DIV
8.4 Pre-trigger
The pre-trigger function serves to record the waveform not
only after but also before triggering has occurred.
Memory Recorder Function
In the memory recorder function, using the recording start
point as 0% and the recording end point as 100%, the
trigger point can be specified in percent. When all trigger
sources are set to OFF, the pre-trigger setting is invalid.
RMS Recorder Function
In the RMS recorder function, using the recording start
point as 0 DIV, the trigger point can be specified in DIV
(0, 5, 10 DIV). When all trigger sources are set to OFF, the
pre-trigger setting is invalid.