276
_____________________________________________________________________
14.4 Recorder Function
______________________________________________________________
Time axis
100*
1
, 200*
1
,500*
1
ms/DIV, 1*
1,3
, 2, 5, 10, 30 s/DIV,
1, 2, 5, 10, 30 min/DIV, 1 h/DIV
Time axis resolution
80 points/DIV (with the printer) *
2
Sampling period
2.5 μs fixed
Recording length
20, 50, 100, 200, 400 DIV, CONT
Display/print layout
Single, dual, quad, X-Y single
Waveform
magnification/
compression
Time axis: x1, x1/2, x1/5, x1/10, x1/20, x1/50,
x1/100, x1/200, x1/500
Voltage axis: x10, x5, x2, x1, x1/2
Storage capacity
Last 400 divisions of data saved in memory
Can be checked by reverse scrolling and reprinted*
2
Auto-print*
2
ON/OFF, real-time print (prints simultaneously as
display)
Logging function*
2
Records measured data as digital values
Interpolation
Line only
X channel
Any channel 1 or 2 (8807-01), 1 to 4 (8808-01)
Y channel
Any channel other than the X channel
Available recording
area
100 mm
x
100 mm (3.94"
x
3.94")
(10 DIV x 10 DIV)
X-Y axis resolution
20 dots/DIV (with the display)
80 dots/DIV (horizontally x vertically) (with the
printer*
2
)
X-Y Sampling period
Dot display: 250 μs fixed,
Line display: 500 μsmax
Recording length
CONT only
Interpolation
Line/Dot only
14.4 Recorder Function
T-Y waveform
X-Y waveform
(*1): screen display only /(*2): when the 8992 is connected.
(*3): when the 8992 is connected.(can be printed only when using the AC
adapter)