EasyManua.ls Logo

Hioki 8808-51 - Trigger Source AND;OR Linking

Hioki 8808-51
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144
_____________________________________________________________________
8.6 Trigger Source AND/OR Linking
______________________________________________________________
Flashing cursor
OR
Triggering is activated when any one of the trigger
conditions is met.
AND
Triggering is activated when all of the trigger
conditions are met.
0V
OR
A
ND
NOTE
If the trigger source is set to AND, and the trigger source setting
conditions have already been met when the START button is
pressed, triggering does not occur. When the conditions are not
met once but met subsequently, triggering occurs.
8.6 Trigger Source AND/OR Linking
The analog trigger, logic trigger, external trigger, and timer
trigger can be linked with the AND/OR logical operators.
Procedure
Trigger Screen
1. Press the
SET
button to display the Trigger Screen.
2. Move the flashing cursor to
Source
and
select the trigger source using the
buttons.
When the level trigger setting is 0.00 V(
) for channels 1
and 2, the following difference between the effect of
AND/OR linking is shown.

Table of Contents