278
_____________________________________________________________________
14.6 Harmonic Wave Analysis Function
______________________________________________________________
Object of
measurement
Select from among independent between channels
(UNIQUE), one-phase two-wire (2 systems), one-phase
three-wire, or three-phase three-wire.
Input settings
UNIQUE
Other settings
Select from among "voltage," "9018," "9132," "3283,"
"3284," "3285," and "9322."
It is possible to set odd channels to voltage and even
channels to current. (When measuring power, it is
necessary either to select input type "9018" or "9132," or
else to select input type "voltage" and then use a clamp-
on probe with good phase precision and apply current
scaling.)
Measurement range
1. Voltage : 5, 10, 20, 50, 100, 200, 500 mVr/DIV
1, 2, 5, 10, 20, 50 Vr/DIV
2. 9018 : 10, 20, 50, 100, 200, 500 Ar
3. 9132 : 20, 50, 100, 200, 500, 1000 Ar
4. 3283 : 10 m, 100 m, 1, 10, 200 Ar
5. 3284 : 20, 200 Ar
6. 3285 : 200, 2000 Ar
7. 9322 : 50, 100 Vr
Base frequency
range
45 to 65 Hz
Automatic setting or manual setting (0.1 Hz resolution)
Analysis degree
Base wave to 40th degree
Measurement precision (on the x1 display)
(Base wave to 20th degree)
(21st degree to 40th degree)
(1.5%rdg. + 15 dgt.)
(3%rdg. + 15 dgt.)
Phase precision ("a" is the component ratio
of the harmonic wave across the full span.)
(Base wave to 20th degree)
(21st degree to 40th degree)
(3.00 + 2.89a + 1/50a)
(6.74 + 7.80a + 1/75a)
Sampling speed
400 ks/s
Number of FFT
operations
512 points (sample extraction for one cycle of the base
wave)
14.6 Harmonic Wave Analysis Function
14.6.1 General Specifications