2.7 Connecting the 9018-10, 9132-10 CLAMP ON PROBE
______________________________________________________________
DANGE
R
Always connect the clamp-on probe to the secondary
side of a breaker. On the secondary side of a breaker,
even if the lines are shorted the breaker can trip and
prevent an accident. On the primary side, however, the
current capacity may be large, and in the event of a
short-circuit there may be a serious accident.
Before using the 9018-10, 9132-10 CLAMP ON PROBE,
be sure to carefully read its instruction manual and
familiarize yourself with the operation principles of this
product.
9018-10
Source
9132-10
Load
Load
Source
NOTE
To measure current, set the input coupling of the product to
"CLAMP." For details, refer to Section 7.3.4.
When the clamp-on probe is used for measurement, the
measurement precision will be affected both by the product
precision and clamp-on probe precision ratings.
When using the 9199 CONVERSION ADAPTOR, the 9018,
9132 CLAMP ON PROBEs can be used.
When measuring electric current, make sure that the direction of
the CLAMP ON PROBE matches the direction of electric
current.The direction of connection for the 9132-10 is opposite
from the 9018-10. If the CLAMP ON PROBE is faced in the
opposite direction, the phase will shift 180 degrees from the
actual value.
2.7 Connecting the 9018-10, 9132-10
CLAMP ON PROBE
This product measures input voltage, so although current
cannot be measured directly, current can be measured by
the voltage output from the HIOKI 9018-10, 9132-10
CLAMP ON PROBEs.
Connect the BNC connector of the CLAMP ON PROBE to
the analog input terminal. For details, refer to Section 2.4.