8.8 Internal Trigger
______________________________________________________________
Flashing cursor
Flashing cursor
Sine 50 Hz
Compare with 50-Hz sine wave.
Sine 60 Hz
Compare with 60-Hz sine wave.
Prev 50 Hz
Compare with 50-Hz previous cycle.
Prev 60 Hz
Compare with 60-Hz previous cycle.
Procedure
Trigger Screen
1. Press the
SET
button to display the Trigger Screen.
2. Move the flashing cursor to the position shown in the
figure below and select
JUDGE
using the buttons.
An error results if the measurement starts when the Time
axis setting is 20 ms or more.
3. Move the flashing cursor to the position shown in the
figure below and select the reference waveform using the
buttons.