EasyManua.ls Logo

Hioki 8808-51 - 10.1 Environment Screen (SYSTEM)

Hioki 8808-51
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
162
_____________________________________________________________________
8.8 Internal Trigger
______________________________________________________________
LA1 1
LA2 0
LA3
X
LA4
X
Trigger pattern
OR
The condition LA1
is met so that
triggering occurs.
The condition LA2 is met,
but LA1 has already met,
so that triggering does not
occurs.
AND
Trigger pattern
LA1 1
LA2 0
LA3 X
LA4
X
NOTE
If the conditions are met already when measurement is started
(AND: all trigger patterns are met, OR: one trigger pattern is
met), triggering does not occur. Triggering only occurs if the
conditions are removed and then met again. Refer to the figure
(1).
The trigger source "AND" cannot be set between the logic
trigger and analog trigger of rms function.
Setting example of the Logic Trigger
(1) If the trigger pattern has been set to "10XX" with the
operator OR, then triggering occurs as shown in the figure
below.
(2) If the trigger pattern has been set to "10XX" with the
operator AND, then triggering occurs as shown in the figure
below.

Table of Contents

Related product manuals