ix
_____________________________________________________________________
Notes on Use
______________________________________________________________
ARNING
When using the batteries, use LR6 alkaline batteries
or 9447 BATTERY PACK. In this case, do not use
new batteries with old ones. Also, do not mix
batteries of different types. We take no responsibility
for an accident using the batteries and battery packs
which are not specified.
The logic inputs are not floating. Although two sets
of logic probes can be connected, they all have a
common ground with the main product.
CAUTION
Use only the specified input cord. Using a non-specified
cable may result in incorrect measurements due to poor
connection or other reasons.
To avoid damage to the product, do not input a
voltage/current exceeding the rated maximum to the
external input terminals.
This product should be installed and operated indoors only,
between 5 and 40
and 35 to 80%RH.
Do not store or use the product where it could be exposed
to direct sunlight, high temperature or humidity, or
condensation. Under such conditions, the product may be
damaged and insulation may deteriorate so that it no longer
meets specifications.
Do not use the product where it may be exposed to
corrosive or combustible gases. The product may be
damaged.
This product is not designed to be entirely water- or dust-
proof. To avoid damage, do not use it in a wet or dusty
environment.
Preliminary Checks
Before using the product the first time, verify that it
operates normally to ensure that the no damage occurred
during storage or shipping. If you find any damage, contact