8.11 Trigger Output Terminal
______________________________________________________________
8808-01
(
A
)
8808-01
(
B
)
8808-01
(
C
)
NOTE
Using the auto ranging function causes a trigger signal to be
output. Care is therefore required when using the auto ranging
function when the trigger output terminal is in use. (Only a
memory recorder function)
A trigger output signal is also generated when a trigger is
applied while "AUTO" is selected as the trigger model.
Example of synchronous measurement using
multiple 8808-01 products
For parallel synchronous measurement, connect 8808-01(A)
trigger output to 8808-01(B) and (C) external trigger inputs
as shown in the figure. External triggering should be
enabled (ON) in the 8808-01(B) and (C) products to
receive the external trigger.
When the trigger condition is set on 8808-01(A) and
enabled, trigger pulses are output from the TRIG OUT
terminal. The 8808-01(B) and (C) products receive these
pulses and apply the trigger to start measurement.