9.4 DMM Function
______________________________________________________________
Measuring object
Commercial mains (50/60 Hz, auto-
selection) and DC signals
Displays
RMS value, instantaneous value
(corresponds to scaling value)
Update rate
1s
Sampling rate
4kS/s
Number of displayed
digits
4 (The lowest digit is displayed as 0 for
values 0 to 4, and as 5 for values 5 to 9.
The actual digit is displayed while setting
scaling)
Accuracy
3% rdg. 5 dgt.
NOTE
The maximum displayable numeric value is 5499. The display
changes to the next lower range when the count is less than 500.
The minimum digit display is either 0 or 5.
When current measurement, auto-ranging is not available. To
change the range, set on the Channel Screen. The setting of
current range should be same as clamp range.
The displayed color of the digital values for each channel
corresponds to the waveform display color set in 7.3.1.
Specifications