4.8 Extended Status Screen Setup (MEM)
______________________________________________________________
START
END
Select ON/OFF
Extended Status Screen
Display the Extended Status Screen.
Select whether to enable Sequential Save.
Set the per-block recording length.
Set the number of memory blocks.
Select whether waveforms should be
displayed on screen after recording to each
block.
Select which blocks to display on screen
after measurement.
Select the block at which to start
measurement.
Select the block at which to end
measurement.
Press the SET button to return to the Status
Screen.
Set the Recording Length
Set the Number of Divisions
Select Subsequent
Waveform Display
Select the Display Block
Select the Starting Block
Select the End Block
NOTE
Dead time during which no further data is sampled, in order to
permit display and recording after the acquired data is recorded
to one block, is about 4 ms.
Dead time, however, may slightly fluctuate by key operation.
While the roll mode is being used in memory recorder function
(on the Extended Status Screen), the sequential save function is
disabled.
Sequential Save Setting Procedure