Appendix 2 Measurement Data Saving Format
A4
About Status Data
The status data indicates the measurement state when the data was saved, and is displayed as a 32-bit
hexadecimal digit, as follows.
HMx : Invalid harmonic parameter (no harmonic sync)
MRx : Motor analysis options A and/or B over range
MPx : Motor analysis options A and/or B peak over
ULM : Motor analysis options A and/or B sync unlocked
UDP: Display impossible (for example, when the measurement data is clearly invalid immediately after the range is
changed)
UCU : Calculation Impossible (e.g., measurement data is invalid immediately after changing ranges)
HUL : Harmonic sync unlocked
ULx : Channel x sync unlocked
RIx : Channel x current over range
RUx : Channel x voltage over range
PIx : Channel x current peak over
PUx : Channel x voltage peak over
(x is a channel number)
Measurement Value Data Format
bit 31 bit 30 bit 29 bit 28 bit 27 bit 26 bit 25 bit 24
HM4 HM3 HM2 HM1 MRB MRA MPB MPA
bit 23 bit 22 bit 21 bit 20 bit 19 bit 18 bit 17 bit 16
ULM UDP UCU HUL UL4 UL3 UL2 UL1
bit 15 bit 14 bit 13 bit 12 bit 11 bit 10 bit 9 bit 8
RI4 RI3 RI2 RI1 RU4 RU3 RU2 RU1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
PI4PI3PI2PI1PU4PU3PU2PU1
General Measurement
Values
±E±
6-digit decimal mantissa and 2-digit exponent ("+" sign and leading zero are omitted for mantissa.)
Integration Value
±E±
7-digit decimal mantissa and 2-digit exponent ("+" sign and leading zero are omitted for mantissa.)
Time
YYYY/MM/DD //
HH:MM:SS ::
Elapsed Time ::
Elapsed Time (ms) 
Error state Input out of range +9999.9E+99