9.7 Message Reference
131
9
(5) Comparator
Query Comparator results
Syntax
Query
Response
:CALCulate:STATistics:LIMit?
<Hi count (NR1)>,<IN count (NR1)>,<Lo count (NR1)>,<Measurement fault count
(NR1)>,<Out-of-range count (NR1)>
Example
:CALC:STAT:LIM?
1516,9310,737,16,5
Query Standard Deviation
Syntax
Query
Response
:CALCulate:STATistics:DEViation?
<n (NR3)>,<n-1 (NR3)>
Example
:CALC:STAT:DEV?
0.0159E-3,0.0161E-3
Query the Process Capability Indices
Syntax
Query
Response
:CALCulate:STATistics:CP?
<Cp (NR2)>,<CpK (NR2)>
Example
:CALC:STAT:CP?
0.86,0.14
Set and Query Comparator Settings
When making comparator settings by commands, the measurement range is not automatically selected.
Comparator State
Syntax
Command
:CALCulate:LIMit:STATe <1/ 0/ ON/ OFF>
Query
Response
:CALCulate:LIMit:STATe?
<ON/ OFF>
Example
:CALC:LIM:STAT ON
Beeper State
Syntax
Command
:CALCulate:LIMit:BEEPer <OFF/ HI/ LO/ HL/ IN>
Query
Response
:CALCulate:LIMit:BEEPer?
<OFF/ HI/ LO/ HL/ IN>
Example
:CALC:LIM:BEEP HL