6.2 Store as soon as Measurement is Stable (Auto-Memory Function)
73
6
Setting the Number of Values to Store
Acquiring Measured Values Automatically
1
Open the Auto-Memory Settings screen.
2
Enter the number of values to store.
To abort the setting process, press . Settings are abandoned and the display returns to the pre-
vious screen.
3
Accept the settings and return to the Measurement screen.
Displays the Auto-Memory setting
screen
Total Count Pass Count (IN) Fail Count (Hi/Lo)
Setting range: 1 to 99
To Reset Numerical Values
Deletes entered digits.
This key is enabled only when enter-
ing numerical values.
(Example: 20 values set to be stored)
1
Momentarily disconnect (open-circuit) the probes.
2
Connect the probes to the DUT.
When the measurement is stable, the value is automatically stored and the count is incremented.
When the count reaches the specified number of values, a long beep sounds, and subsequent mea-
surements are not stored.
The (one) last acquired value can be deleted (Undo function (p. 76)).