2.6 Calibration of viscometer
2.6.1 Calibrating viscometer
The calibration of viscometer is to determine the “coefficient” inherent in the
equipment.
Calibrate the viscometer when installing equipment. Normally, calibrating it
once is enough. However, calibrate the viscometer again when changing the
type of ink, and also calibrate a new viscometer after replacement.
Be sure to calibrate the viscometer when beginning to use new ink.
Calibrating the viscometer can be done regardless of the ambient temperature.
Turn on the power.
Press viscometer calibration and Execute at the second page of
circulation control on the user maintenance screen.
First page of viscometer display screen
Second page of viscometer display screen
Press Start calibration .
“Caution” When the operation confirmation of the viscometer is performed
without the calibration, press Operation test.
InkdrainageInkcirculation
Inkstream
adjustment
Pressurerelief
Viscometer
calibration
Circulationenv.
setup
Inkrefill
Makeupinkrefill
Operatingtime
management
N
ext
dis-
p
al
y
Execute
Circulationcontrol
Manual
Shut
down
Cancel
Status:XXXXX
Next
menu
Back
Viscometercalibration
Start
calibration
Operation
test
Status:XXXXX
<thepreviouscalibration> thelatestautoreplenishment
Exec.time 000.00.0000:00 Exec.time 000.00.0000:00
Standard 0.000(mPas
Coefficient 0000
Inktemperature 00(℃)
Standard 0.000(mPas
Coefficient 0000
Inktemperature00(℃)
Inkviscosity 000
anewinkfirst,thenpress<Startcalibration>.
After"inkdrainage"iscompleted,perform"inkrefilling"of
Inkrefillinglog 2004.07.0712:20
Next
menu
Back
Viscometercalibration
Start
calibration
Operation
test
Status:XXXXX
<thepreviouscalibration> thelatestautoreplenishment
Plungerfallingtime
№0.00(sec
Average 0.00(sec
№0.00(sec
№0.00(sec
№0.00(sec
№0.00(sec
Plungerfallingtime
№0.00(sec
Average 0.00(sec
№0.00(sec
№0.00(sec
№0.00(sec
№0.00(sec
Viscometercalibration
Abort
Status:XXXXX
<thepreviouscalibration>
Exec.time 000.00.0000:00
Standard 0.000(mPas
Coefficient 0000
Inktemperature 00(℃)
Proc.status:Inprogress.