Affirm Breast Biopsy Guidance System User Guide
Appendix C: CNR Correction for Biopsy
MAN-06411-002 Revision 003 Page 93
Appendix C CNR Correction for Biopsy
C.1 CNR Correction for Stereotactic 2D Biopsy
Note
System default setting is AEC Table 0 for imaging under stereotactic 2D biopsy mode.
C.1.1 AEC Table 0 (Standard Stereotactic 2D Biopsy Dose)
Prior to Software
After Software Rev 1.7.x
Compression
All Detectors Detector Serial #:
Detector Serial #:
2.0 cm 0.91 0.91 0.91
4.0 cm 1.00 1.00 1.00
6.0 cm 1.27 1.27 1.32
8.0 cm 1.76 1.76 1.88
C.2 CNR Correction for Biopsy Under Tomosynthesis Option
Note
System default setting is AEC Table 0 (Tomosynthesis Option) for imaging under
biopsy mode using digital breast tomosynthesis.
C.2.1 AEC Table 0 (Tomosynthesis Option: Standard Tomo Dose)
Prior to Software
Rev 1.7.x
Detector Serial #:
XX6xxxxx
Detector Serial #:
XX8xxxxx
2.0 cm 0.70 0.70 0.70
4.0 cm 0.91 0.91 0.91
6.0 cm 1.46 1.46 1.55
8.0 cm 2.37 2.37 2.78