Affirm Breast Biopsy Guidance System User Guide
Chapter 7: Care and Cleaning
Page 82 MAN-06411-002 Revision 003
7.4 For General Cleaning
Use a lint-free cloth or pad and apply a diluted dishwashing liquid.
Caution:
Use the least possible amount of cleaning fluids. The fluids must not flow or run.
If more than soap and water is required, Hologic recommends any one of the following:
10% chlorine bleach solution and water with one part commercially available
chlorine bleach solution (normally 5.25% chlorine and 94.75% water) and nine parts
water. Mix this solution daily for best results.
Commercially available isopropyl alcohol solution (70% isopropyl alcohol by
volume, not diluted)
3% maximum concentration of hydrogen peroxide solution
After you apply any of the above solutions, use a pad and apply a diluted dishwashing
liquid to clean any parts that touch the patient.
Warning:
If a paddle touches possible infectious materials, contact your Infection
Control Representative to remove contamination from the paddle.
Caution:
To prevent damage to the electronic components, do not use disinfectant sprays on
the system.
7.4.1 How to Clean the Biopsy Control Module Screen
There are many commercially available products to clean LCD screens. Make sure the
product that you select is free of strong chemicals, abrasives, bleach, and detergents that
contain fluorides, ammonia, and alcohol. Follow the directions of the manufacturer of the
product.