EasyManua.ls Logo

Hologic Affirm - Module Specifications

Hologic Affirm
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Affirm Breast Biopsy Guidance System User Guide
Appendix A: System Specifications
MAN-06411-002 Revision 003 Page 87
A.3 Biopsy Guidance Module
Accuracy of Biopsy Guidance
Module
maximum deviation: 1 mm in either direction of target coordinate
Combined Accuracy of Biopsy
Guidance Module and Biopsy
Device
maximum deviation: 2 mm in either direction of target coordinate
Range of Movement
X-axis: ± 35 mm
Y-axis: + 72.8 mm
Z-axis: + 161 mm
Speed of Motorized Movements
Continuous: No faster than 5 mm per second
Incremental: 0.5 mm steps
Power System
Input from Selenia Dimensions system or 3Dimensions system:
+ 15 Vdc ± 10% and + 5 Vdc ± 10%
Output: + 12 Vdc
A.4 Lateral Arm
Accuracy of Biopsy Guidance
Module and Lateral Arm
maximum deviation: 1 mm in either direction of target coordinate
Combined Accuracy of Biopsy
Guidance Module, Lateral Arm,
and Biopsy Device
maximum deviation: 2 mm in either direction of target coordinate
Range of Movement
Lateral X-axis: 22 cm
A.5 Biopsy Control Module
Display Window
Touch screen controls
Weight
3 pounds
Power System
Input from Biopsy Guidance Module: + 5 Vdc ± 10%

Table of Contents

Other manuals for Hologic Affirm

Related product manuals