Affirm Breast Biopsy Guidance System User Guide
Chapter 5: Biopsy
MAN-06411-002 Revision 003 Page 65
5.5.1 Verify the Position of the Biopsy Device
If desired, use the following steps to verify the position of the biopsy device.
1. Acquire the pre-fire images as necessary to identify the correct needle position.
• Verify the needle position.
• Make adjustments as necessary.
2. If applicable, fire the biopsy device.
3. If desired, acquire the post-fire images.
• Verify the needle position.
• If necessary, make adjustments.
4. If desired, acquire specimens with the attached biopsy device.
5. If desired, acquire post procedure images.
5.5.2 Project Targets on Post-Fire Scout Image
To project targets from the pre-fire tomosynthesis target (scout) onto the post-fire
tomosynthesis target (scout), follow these steps:
1. Select the pre-fire tomosynthesis target (scout) thumbnail. The image shows in the
bottom half of a 2-Up screen on the Image Display monitor.
2. Select the post-fire tomosynthesis target (scout) thumbnail. The image shows in the
bottom half of the 2-Up screen.
3. Select the Project Target button in the Biopsy Options area to show the pre-fire
targets on the post-fire tomosynthesis target (scout).