Affirm Breast Biopsy Guidance System User Guide
Chapter 7: Care and Cleaning
MAN-06411-002 Revision 003 Page 81
7.3 Service Preventive Maintenance Schedule
Table 10: Service Engineer Preventive Maintenance Schedule
Maintenance Task Description
Each Visit
6 Months
Annually
Review user concerns or issues.
Inspect all system components for wear and damage.
Check missing, damaged, or loose hardware. Be sure
covers fit tightly.
Inspect paddles for damage and defects. Replace as
necessary.
Inspect cables and connections for wear and damage.
Inspect all ground wires for damage and correct
connection.
Inspect and clean the unit.
Verify the calibration of the ±15 degree stereo position.
Verify the x-ray field for Biopsy Mode.
Verify all of the Affirm system functions.
Check compression lockout operation.
Verify the Affirm system alignment to the C-arm.
Verify the Needle Guide alignment.
Verify the Affirm system locks in position on the C-arm.
Verify the ID sensors correctly identify the Affirm system
when installed.
Install software/ firmware upgrades if required and if
under warranty/contract.
Perform QAS Procedures to confirm target accuracy.
Complete all required documentation, Preventive
Maintenance, and Service Requests. Make an appointment
for repair service if required.