Affirm Breast Biopsy Guidance System User Guide
Chapter 6: Quality Control
MAN-06411-002 Revision 003 Page 75
6.2.2 QAS Test for Lateral Needle Approach
1. Make sure the Affirm system is installed correctly and all paddles are removed.
2. On the Acquisition Workstation, navigate to the Admin screen.
3. Select the Lateral QAS button.
Figure 71: Lateral QAS Button on the Admin Screen