EasyManua.ls Logo

Hologic ThinPrep 5000 - Clean System; Reports and Logs

Hologic ThinPrep 5000
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
6.65
6
USER INTERFACE
Clean System
This is described in Chapter 8, Maintenance.
Reports And Logs
Figure 6-78 Reports and Logs Button
The Reports and Logs interface presents system information in five forms:
System Events - a log of all system errors excluding UPS power status events or sample
preparation errors that do not interfere with the operation of the instrument. The record of
errors is retained for three years; errors older than three years are purged.
Batch Reports - displays the success or failure of sample processing for each vial tray
processed.
Usage Details - indicates the number of slides successfully created to date, by sequence type.
Carousel Report - displays the status of sample errors resulting in vial or slides left in the
carousel at the end of a batch.
Vial Event Report - displays a detailed list of all sample processing errors that occurred in the
most recent batch.