MAINTENANCE
8.2
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
8
DAILY
Change Fixative Reagent
The fixative alcohol in any bath should be changed out every 100 slides, or daily, whichever comes
first. Consider how your laboratory uses baths in the count to 100. For example, one bath used with
20 slides for 5 batches needs the fixative alcohol changed before the next batch is run (or daily).
Dispose of fix reagents according to your laboratory’s protocols
Clean the fixative bath containers, covers and staining racks according to your laboratory’s
protocols
Wipe Down Slide Cassette
Before filling a slide cassette, wipe down the inside using a dry, non-abrasive cloth. This will keep
down glass dust and help the slides to be removed more easily.
If you notice that slides are sticking together or pick errors occur, make sure the cassettes are clean.
WEEKLY CLEANING
Slide Cassette Area
Figure 8-1 Slide Cassette Area
Open the door to the slide cassette compartment. Remove slide cassettes that are present and clean
around the area using deionized water and lint-free cloths.