ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
8.1
8
MAINTENANCE
Chapter Eight
Maintenance
Table 8.1: Routine Maintenance
Daily or
more
Change fixative every 100 slides or daily,
whichever comes first
Before filling a slide cassette, wipe down the
inside with a dry, non-abrasive cloth
Weekly
Clean around the slide cassettes, the carousel
and dispersion area
Clean slide holder pneumatic suction cups at
the slide cassette area and the processing area
Check and empty the filter waste bin and the
slide waste bin
As needed
Empty waste bottle
Clean the touch screen
Check and empty material left in the robot
arm elevator shaft
Clean input carousel and dust cover
Change absorbent pads
Remove and clean drip trays
Change the fume extractor carbon filter yearly
or as desired
Change the fume extractor HEPA filter when
prompted by the instrument