USER INTERFACE
6.70
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
6
Figure 6-86 Batch Report Example
To print a report, press the
Print
button (if your processor is configured with a printer).
To save a report as a text file, press the
Save to USB
button. See the next section.
To close a report, press the
Done
button.
Note:
The system will retain batch reports for eight weeks and then purge them from the database.
Should your lab require longer record retention, plan to print or download the batch reports.
Find Vial by ID
Figure 6-87 Find Vial by ID Button
The batch report list screen has a feature that enables searching for a vial ID using all or some of the
accession number. Press the
Find Vial by ID
button.
ThinPrep® 5000 Batch Report
Start Time:
End Time:
Lab:
Instrument:
Serial number:
Sequence:
Status:
10/21/2010 10:15 AM
10/21/2010 11:45 AM
Hologic
T5000
D002K09DP
Gyn
OK
38 Vials Processed: 33 OK 5 Events
83668809999025
79000151115002
08387390999138
83805969999060
10019939999083
10019979999206
83668 2999923
1-1
1-3
1-4
1-5
1-6
1-7
-
Sample is dilute
-
Sample is dilute
Sample is dilute
-
OK
5000
OK
5000
5000
OK
O
Tray-Pos.
Vial/Slide ID
DescriptionStatus
2 Sample Errors
Tray-Pos.
Carousel Pos.
Vial/Slide ID Status Description
83668909999150
79000781178110
2
1
5010
5002
1-2
1-8
Insufficient fluid or no filter present
Failed to uncap vial
Batch report header with
identifying information and
batch status
List of sample errors (no slide
was made)
List of vials processed and
any events (Slide was made
but needs operator
attention.)