EasyManua.ls Logo

Hologic ThinPrep 5000 - Page 211

Hologic ThinPrep 5000
346 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
6.77
6
USER INTERFACE
Figure 6-96 Carousel Report
ThinPrep® 5000 Carousel Report
Start Time:
End Time:
Lab:
Instrument:
Serial number:
Autoloader Serial Number:
10/21/2010 10:15 AM
10/21/2010 11:45 AM
Hologic
T5000
D002K09DP
D004M10DA
Carousel Report Summary
Pos.
Vial Filter Slide
Tray-Pos. Vial Status
1-1
-
1-14
2-16
2-21
-
3-4
4-13
-
4-40
-
-
-
-
-
-
-
-
-
-
Yes
No
Yes
Yes
Yes
No
Yes
Yes
No
Yes
No
No
No
No
No
No
No
No
No
No
No
No
Yes
Yes
No
Yes
No
No
No
Yes
No
No
No
No
No
No
No
No
No
No
Yes
No
No
Yes
No
No
Yes
Yes
No
Yes
No
No
No
No
No
No
No
No
No
Yes
Error Hold
-
Unprocessed
Error Hold
Unprocessed
-
Waiting Return
Waiting Return
-
-
-
-
-
-
-
-
-
-
-
-
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Carousel Position 1 Details:
Yes
No
Yes
1-1
Gyn
Error Hold
74041829999
74041829999
Gyn
Vial Status:
Vial ID:
Slide ID:
Slide Type:
Vial Present:
Filter Present:
Slide Present:
Tray-Pos:
Vial Type:
Status Details:
5010 - Insufficient fluid or no filter present
Carousel Position Status Map
1 3456 78 102 9 11 12 13 14 15 16 17 18 19 20
Position Status
NOTE: Empty Position = X
Report header contains date/time
stamp and instrument identification
Carousel Report Summary indicates the
presence of media in each position where
the vial was originally located (vial tray-
position) a brief vial status
Carousel Position Details indicate
Media present
Error type
Accession ID and slide type
Carousel Position Status Map
provides a quick visual overview of
where material is present in the
carousel.

Table of Contents

Other manuals for Hologic ThinPrep 5000

Related product manuals