INSTALLATION
2.8
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
2
Extended Shutdown
If the instrument is to be shut down for an extended amount of time, or be taken out of service,
empty the waste bottle (Maintenance chapter), remove any items that may be on board and close all
doors. Follow the instructions for Normal Shutdown. Completely remove power to the instrument
by unplugging the power cord from the UPS and the UPS from the outlet.
WARNING
: Never disconnect the UPS wall plug when the instrument is running on battery
power. The instrument needs to remain connected to ground through the UPS.