HP 8114A Service Guide
-222
-222
-222
-222
-222
-222
-222
-222
Data out of range
HP-IB message = “Width
(%)
> Width (%) Limit,”
Local message = “Width
(%)
>
Width
(%)
Limit”
Description:
Width in percent is greater than allowed user limit
Data out of range
HP-IB message = “OUTPUT: High Level
>
High-Amp Limit”
Local message = “OUTPUT: High Level
>
High-Amp Limit”
Data out of range
HP-IB message = “OUTPUT: High Current
>
Low
Current”
Local message = “OUTPUT: High Current
>
Low
Current”
Data out of range
HP-IB message = “OUTPUT: High Level (V) close
abs. limit”
Local message = “OUTPUT: High Level (V) close
abs. limit”
Data out of range
HP-IB message = “OUTPUT: Low Level (V) close
abs. limit”
Local message = “OUTPUT: Low Level (V) close
abs. limit”
Data out of range
HP-IB message = “OUTPUT: High Level (A) close
abs. limit”
Local message = “OUTPUT: High Level (A) close
abs. limit”
Data out of range
HP-IB message = “OUTPUT: Low Level (A) close
abs. limit”
Local message = “OUTPUT: Low Level (A) close
abs. limit”
Data out of range
HP-IB message = “OUTPUT: Amplitude (V) close
high limit”
Local message = “OUTPUT: Amplitude (V) close
high limit”
Troubleshooting
5-29