A) Amplitude
Test
(from
509
into
503)
Procedure
1.
Connect HP
8114A
t,o
the DVM as shown:
‘N
TVE
TO BNC ADAPTOR
2. On the HP 8114A set up the
UUTPUT
page as shown in the
following illustration:
Configuring the Output Rage
3. Set HP 8114A Delay to
0
ns
4.
‘kke
the baseline reading as
B,,a
5. Set HP 8114A Delay to 50 ms and Amplitude to 1 V
6.
‘E&e
the high-level reading as
H,,Q
7. Calculate the Amplitude:
A
5062
=
(H5061
-
B5od
x
Gl
8. Repeat steps 3 to 7 for the following settings:
Amplitude
Amplitude limit TR
Entrj
minimum maximum
1v
0.89
v
1.11
v 6-l
2v
1.38
v
2.12
v 6-2
5V
4.86
v
5.15
v 6-3
10 v
9.80
v
10.2
v 6-4
20 v
19.7
v
20.3 V
6-5
60
v
49.4
v
50.6 V
6-6
‘Ihble
3-10. Amplitude Levels: 503 into
5OQ
7
J
Testing the HP
8114A
3.31