EasyManua.ls Logo

HP 8114A - Variable Baseline Option 001

HP 8114A
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP
8114A Specifications
Variable Baseline Option 001
L
--
1 Maximum Duty Cycle, see Figure 6.1
BZj
Maximum Duty Cycle 15%
Figure 2-2. Baseline Duty-cycle
/
Amplitude Ranges
Refer to Figure 2-2. Under the following conditions Figure 2-l still
applies for the maximum duty-cycle:
n Positive pulse with negative Baseline
m
Negative pulse with positive Baseline
n -5 V <Baseline
i:
+ 5
V,
negative or positive pulses
Under the following conditions maximum duty-cycle is 15%:
m
Baseline
>
+ 5 V and positive pulses
m
Baseline
c-5
V and negative pulses
Note also, that the pulse is limited to a
SO
V window (500 into 5061)
so that for positive pulses with positive Baseline, or negative pulses
with negative baseline, the maximum available amplitude becomes
limited by the Baseline setting
2.8
HP 8114A Pulse Generator

Table of Contents

Related product manuals