HP 8114A Service Guide
-802
-803
Couldn’t write that key code
HP-IB message = “Couldn’t execute Key: [no]
Local message = “Couldn’l execute Key: [no]
Description:
[no] = key number
Demofile
is bigger than read buffer
HP-IB message = “Demo file too big
Local message = “Demo file too big
fl
Register Usage
The HP 8114A firmware uses the Standard Event Status, Operation
Status, and Questionable Status registers for reporting instrument
status, in accordance with the SCPI standard. The following table lists
the bits used in each register, and what they are used for.
lhble
5-13. Bits Used in SCPI Registers
Register
Bit
Description
STANDARD-EVENT-STATUS-REGISTER
0
Set when inst. has completed all pending operations
7
Set to indicate that instrument has powered-up
)PERATION-STATUS-REGISTER
1
Set when inst. starts changing its output signals
1
Cld. when inst. finishes changing output signals
2
Set when inst. starts changing its range
2
Cleared when inst. finishes changing its range
4
Set when inst. starts a frequency measurement
4
Cld. when inst. finishes frequency measurement
QUESTIONABLE-STATUS-REGISTER
0
Set the
QUEStionable
STAT
US
bit for VOLTage
0
Clear the
QUEStionable
STATUS bit for
VOL%ge
1
Set the
QUEStionable
STAT
US
bit for CURRent
1
Clear the
QUEStionable
STATUS bit for
CURRent
2
Set the
QUEStionable
STAT
US
bit for TIME
2
Clear the
QUEStionable
STAT
US
bit for TIME
5
Set the
QUEStionable
STAT
US
bit for VOLTage
5
Clear the
QUEStionable
STAT
US
bit for VOLTage