HP 8114A Service Guide
-330 Selftest
failed
-330
-330
-330
-800
-801
HP-IB message = “Failure in AMPLIFIER circuitry”
Local message = “Failure in AMPLIFIER circuitry”
Description:
The AMPLIFIER, or the OFFSET circuitry may be
defective.
Selftest
failed
HP-IB message = “Failure in AMPLIFIER or
OFFSET circuitry’
Local message = “Failure in AMP/OFFS circuitry”
Description:
The AMPLIFIER or the OFFSET circuitry may be
defective.
Selftest
failed
HP-IB message = “No output signal PERIOD
TIGER”
Local message = “No output signal PERIOD TIGER”
Description:
An output signal cannot be found by the
selftest
following the period circuitry.
Selftest
failed
HP-IB message = “Failure in TRIGGER OUTPUT
circuitry”
Local message = “Failure in TRIGGER OUTPUT
circuitry”
Description:
An output signal cannot be found by the
selftest
for
the TRIGGER OUTPUT AMPLIFIER.
The file Header isn’t right
HP-IB message = “File isn’t a demo file
”
Local message = “File isn’t a demo file
n
There are some unexpected characters
HP-IB message = “SYNTAX ERROR: ROW: [no]
’
Local message = “SYNTAX ERROR: ROW: [no]
”
Description:
[no] = row number
Troubleshooting
5-41