Test 7: Variable Baseline (Option 001)
Note
1
P
This
t,est
is only to be performed if Option 001 is installed.
Test Specifications
Range:
-25Vto
+25V
Accuracy: 51%
klO0
mV
50.5%
of amplitude
500
source impedance only
Equipment Needed
1. Digitizing Voltmeter (DVM)
2. High Power attenuator
3.
508,
O.l%,
1OW
Feedthrough
Procedure
1. On the HP 8114A set up the OUTPUT page as shown in the
following illustration:
Configuring the Output Page
2. Connect HP 8114A to the DVM as shown:
/
iP
8 I I 4A
IINDER
TEST
50
OHM
FEEDTHROUGH
Equipment Set-up 1 for Variable Baseline
Test
edings for the following baseline settings (VB). Multiply the
r
3. Set the DVM HP 3458A to:
Function: DCV
Trigger: TRIG EXT
AD-Converter integration time NPLC: 0.1
(NPLC = Number of Power Line Cycles)
3.34
Testing the HP 8114A