4. Take the DVM readings for the following baseline settings (VB).
Multiply the reading by the attenuation factor
G1
[derived from the
amplitude test, test 6).
B =
VBxG,
Compare the calculated value with the given limits:
l’&ble
3-12. Baseline Levels Test
E
laseline
Leve.
-25 v
-20 v
-10 v
-5 v
-2 v
l o v
+2v
+5v
+ 10 v
+20
v
125
v
T
Li
minimum
-25.355 V
-20.305 V
-10.205 v
-6.155 v
-2.125 V
-0.105 v
+
1.876 v
+ 4.845 V
+9.795 V
*
19.695
\
+
24.645
L
mi
I
ts
maximum
-24.645 V
-19.696 V
-9.795 V
-4.435 v
-1.876 V
+0.106
v
t 2.125 V
+5.166
v
+
10.205 I
+
20.305 I
+
26.355
1
‘R
Entq
7-
1
7-2
7
-
3
7-4
7-5
7-6
7-7
7-K
7-9
7 10
7-
11
I
Testing the HP
8114A
3-35