Test 8: Transition Time
The following tests are required:
1.
From 500 into 500, amplitudes
>5
V
2. From HIZ (High-Z) into
50R
amplitudes
>lO
V
Test Specifications Range:
~7
ns for amplitudes
>5
V
(5061
into 503)
~12
ns for amplitudes
>lO
V (HIZ {High-Z} into
5OQ)
Equipment Needed
Digitizing Oscilloscope with Accessories
High Power Attenuator
3.36
Testing the HP